Reading a Peptide Mass Spectrum: ESI, MALDI, Charge States and Isotopes

Peptide mass spectrometry measures the mass-to-charge ratio (m/z) of ionised peptide molecules, and is the standard way to confirm that a synthetic peptide has the expected molecular mass. Electrospray ionisation (ESI) usually produces a series of multiply charged ions, while MALDI mostly produces singly charged ions. Reading a spectrum means working out the charge state, finding the monoisotopic peak, and comparing the calculated mass with the theoretical mass.

Written by
Team Iconic Peptides · Editorial policy
Last reviewed
11 October 2026
Reading time
10 min
A mass spectrum: isotope clusters at several charge states (illustrative)
A benchtop mass spectrometer beside a computer screen showing a spectrum with a cluster of sharp peaks

In plain English

A mass spectrometer is a very precise scale for molecules. It cannot weigh a molecule directly. First it gives the molecule an electric charge (ionisation), then it measures how the charged molecule moves in an electric or magnetic field. The machine reports mass divided by charge, written m/z [3, 4].

For peptides (short chains of amino acids), this is the main way to check identity. If a peptide should weigh 1,045.53 daltons (Da, the unit used for molecular mass) and the spectrum shows that mass, the identity is supported. If it shows something else, the vial may hold a different molecule. In one study of a research peptide bought from five manufacturers, one product turned out to be a completely different peptide [7].

Mass spectrometry confirms mass. On its own, it does not measure purity or the amount of peptide in a vial.

The short version

  • The spectrum shows mass ÷ charge, not mass.
  • One peptide can appear as several peaks with different charges.
  • Isotopes make each peak a small cluster.
  • Identity needs mass spectrometry; purity needs a separating method such as HPLC.

How peptides become ions: ESI and MALDI

A peptide must carry a charge and be in the gas phase before it can be measured. Two “soft” ionisation methods do this without breaking the peptide apart.

ESI (electrospray ionisation)MALDI (matrix-assisted laser desorption/ionisation)
How it worksA peptide solution is sprayed from a fine needle at high voltage; charged droplets shrink until ions are released [1]The peptide is mixed with a light-absorbing matrix, dried on a plate and hit with a laser pulse [2]
Typical chargeSeveral charge states at once (1+, 2+, 3+ and higher for larger molecules) [1]Mostly singly charged ions ([M+H]⁺) [4]
CouplingConnects directly to liquid chromatography (LC-MS)Usually run as a separate spot-by-spot analysis
Common use for synthetic peptidesIdentity plus impurity identification during LC-MSFast identity checks

Electrospray was shown to give “coherent sequences of peaks” in which neighbouring ions differ by one charge, which is the pattern seen for peptides and proteins today [1]. Laser desorption with a matrix was shown in 1988 to ionise intact proteins above 10,000 Da [2].

What the axes mean

The x-axis is m/z: the mass of the ion divided by its number of charges. The y-axis is relative intensity, scaled so the tallest peak is 100%. Intensity reflects how well a species ionised and was detected, so peak height on a mass spectrum is not a reliable measure of how much of each species is present. That is one reason purity is measured by HPLC rather than MS [3, 9].

Charge states: one peptide, several peaks

In positive-mode ESI, a peptide picks up protons (H⁺, mass 1.00728 Da). An ion with z extra protons appears at:

m/z = (M + z × 1.00728) ÷ z

where M is the neutral monoisotopic mass.

Worked example: angiotensin II

Angiotensin II (sequence DRVYIHPF) is a common calibration peptide. Its formula is C₅₀H₇₁N₁₃O₁₂, with a monoisotopic mass of 1045.5345 Da and an average molecular weight of 1046.2 g/mol [8].

IonCalculationExpected m/z
[M+H]⁺(1045.5345 + 1.00728) ÷ 11046.54
[M+2H]²⁺(1045.5345 + 2.01456) ÷ 2523.77
[M+3H]³⁺(1045.5345 + 3.02184) ÷ 3349.52
[M+Na]⁺1045.5345 + 22.989221068.52

An ESI spectrum of this peptide would typically show the 2+ ion strongly, with smaller 1+ and 3+ peaks. A MALDI spectrum would be dominated by the 1+ ion.

Working out the charge from two neighbouring peaks

If two peaks belong to the same peptide and differ by one charge, the charge of the higher-m/z peak is:

z = (lower m/z − 1.00728) ÷ (higher m/z − lower m/z)

Using 349.52 and 523.77: (349.52 − 1.00728) ÷ (523.77 − 349.52) = 348.51 ÷ 174.25 = 2.0. The 523.77 peak is therefore 2+, and the neutral mass is 2 × (523.77 − 1.00728) = 1045.53 Da, matching the theoretical value [8]. Software performs this “deconvolution” automatically, but the arithmetic is the same.

Isotopes: why each peak is a cluster

Carbon in nature is about 1.1% carbon-13. A peptide with 50 carbon atoms therefore has a good chance of containing at least one ¹³C atom, which adds about 1.003 Da. Each ion appears as an isotope envelope: a main peak followed by smaller peaks.

  • Monoisotopic peak. The first peak in the cluster, containing only the lightest isotopes (¹²C, ¹H, ¹⁴N, ¹⁶O). High-resolution instruments report this value [3].
  • Average mass. The weighted mean over all isotopes. This is the “molecular weight” listed in databases and on certificates (1046.2 for angiotensin II, versus 1045.53 monoisotopic) [8].
  • Spacing reveals charge. Isotope peaks are about 1.003 apart for a 1+ ion, 0.50 apart for 2+, and 0.33 apart for 3+. On a high-resolution spectrum, the spacing alone gives the charge.

As peptides get larger, more of the molecules carry heavy isotopes, and the monoisotopic peak becomes a smaller part of the cluster. Comparing a monoisotopic measurement with an average theoretical mass (or the reverse) is a common source of an apparent “1 Da error”.

Adducts and contaminants

Not every peak is the target peptide or an impurity from synthesis. Common extra peaks include:

Observed shift from [M+H]⁺Likely cause
+21.98Sodium adduct ([M+Na]⁺ instead of [M+H]⁺)
+37.96Potassium adduct ([M+K]⁺)
+15.99Oxidation, often of methionine [6]
+0.98Deamidation of asparagine or glutamine [6]
Loss of one residue massDeletion sequence from synthesis [6]

Sodium and potassium adducts, polymer contaminants and solvent clusters are well-documented background signals in ESI and MALDI [5]. Trifluoroacetic acid (TFA), common in peptide HPLC and as a counter-ion, can suppress the ESI signal through ion pairing and surface-tension effects [10]. See Peptide Counter-Ions: TFA and Acetate and Common Peptide Impurities.

Tandem MS: confirming the sequence

Matching the intact mass shows that the formula is plausible, but different sequences can share the same mass. Tandem mass spectrometry (MS/MS) breaks the selected ion into fragments. Breaks along the backbone produce b ions (containing the N-terminus) and y ions (containing the C-terminus). The mass differences between consecutive fragments correspond to individual amino acid residues, so the sequence can be read from the spectrum [3]. Leucine and isoleucine have identical masses and are not distinguished by standard fragmentation.

Reading the MS section of a COA

A useful mass spectrometry result on a Certificate of Analysis states:

  1. The method (ESI or MALDI; LC-MS or direct infusion).
  2. Observed and theoretical mass, with a note on whether each is monoisotopic or average.
  3. The ion used (for example [M+2H]²⁺ at m/z 523.77) or a deconvoluted neutral mass.
  4. An attached spectrum with labelled peaks.
  5. An acceptance criterion, such as agreement within a stated tolerance.

International guidance on specifications for biotechnological products lists identity testing as a core part of a specification [9]. For how MS fits with HPLC purity, see HPLC, LC-MS and UV-Vis and How to Read a Peptide COA.

What this means in Australia

  • Identity testing does not change regulatory status. A mass spectrum confirms what a molecule is. It does not enter a product on the Australian Register of Therapeutic Goods (ARTG), and unapproved peptide products have not been assessed by the TGA [11].
  • Scheduling depends on the substance. Several peptides are named in the current Poisons Standard (October 2026), some in Schedule 4 (prescription-only) and Appendix D, clause 5 (possession without authority is illegal) [12]. Confirming identity by MS is how a laboratory knows which substance it actually holds.
  • Labels don’t change status. The TGA has stated that a “research use only” disclaimer does not make supply lawful, change regulatory status, permit importation, or remove advertising or supply obligations [11].
  • Accredited testing. In Australia, NATA accredits testing laboratories to ISO/IEC 17025, which is a useful signal when judging who produced a mass spectrometry report [13].

This is general information, not legal advice. Laws change; check the current Poisons Standard and TGA guidance.

Frequently asked questions

What does m/z mean in mass spectrometry?

m/z is the mass of an ion divided by its charge. A peptide carrying two protons appears at roughly half its mass, so the same molecule can produce several peaks [1].

What is the difference between ESI and MALDI?

ESI sprays a solution at high voltage and usually produces multiply charged ions, which suits LC-MS. MALDI uses a laser and a matrix and mostly produces singly charged ions [1, 2, 4].

Why is the measured mass 1 Da different from the expected mass?

The usual cause is comparing a monoisotopic value with an average mass, or picking the wrong peak in the isotope cluster. Checking which mass type each number represents usually resolves it [8].

Can mass spectrometry measure peptide purity?

Not reliably on its own. Ionisation efficiency differs between molecules, so peak heights don’t reflect amounts. Purity is measured by a separating method such as HPLC, often combined with MS as LC-MS [3, 9].

What is an [M+Na]⁺ peak?

It is the peptide carrying a sodium ion instead of a proton. It appears about 22 units above [M+H]⁺ and is a common background signal rather than an impurity in the peptide itself [5].

References

  1. Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989;246(4926):64–71. doi:10.1126/science.2675315. PMID: 2675315
  2. Karas M, Hillenkamp F. Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltons. Anal Chem. 1988;60(20):2299–2301. doi:10.1021/ac00171a028. PMID: 3239801
  3. Steen H, Mann M. The ABC’s (and XYZ’s) of peptide sequencing. Nat Rev Mol Cell Biol. 2004;5(9):699–711. doi:10.1038/nrm1468. PMID: 15340378
  4. Domon B, Aebersold R. Mass spectrometry and protein analysis. Science. 2006;312(5771):212–217. doi:10.1126/science.1124619. PMID: 16614208
  5. Keller BO, Sui J, Young AB, Whittal RM. Interferences and contaminants encountered in modern mass spectrometry. Anal Chim Acta. 2008;627(1):71–81. doi:10.1016/j.aca.2008.04.043. PMID: 18790129
  6. D’Hondt M, Bracke N, Taevernier L, et al. Related impurities in peptide medicines. J Pharm Biomed Anal. 2014;101:2–30. doi:10.1016/j.jpba.2014.06.012. PMID: 25044089
  7. De Spiegeleer B, Vergote V, Pezeshki A, Peremans K, Burvenich C. Impurity profiling quality control testing of synthetic peptides using liquid chromatography-photodiode array-fluorescence and liquid chromatography-electrospray ionization-mass spectrometry: the obestatin case. Anal Biochem. 2008;376(2):229–234. doi:10.1016/j.ab.2008.02.014. PMID: 18342612
  8. National Center for Biotechnology Information. PubChem Compound Summary for CID 172198, Angiotensin II. pubchem.ncbi.nlm.nih.gov/compound/172198
  9. International Council for Harmonisation. ICH Q6B: Specifications: test procedures and acceptance criteria for biotechnological/biological products. database.ich.org
  10. Apffel A, Fischer S, Goldberg G, Goodley PC, Kuhlmann FE. Enhanced sensitivity for peptide mapping with electrospray liquid chromatography-mass spectrometry in the presence of signal suppression due to trifluoroacetic acid-containing mobile phases. J Chromatogr A. 1995;712(1):177–190. doi:10.1016/0021-9673(95)00175-m. PMID: 8556150
  11. Therapeutic Goods Administration. Understanding your responsibilities when importing, compounding and supplying unapproved peptide products. Safety advisory, 13 April 2026. tga.gov.au. Accessed 11 October 2026.
  12. Therapeutic Goods (Poisons Standard—October 2026) Instrument 2026 (F2026L01327). Federal Register of Legislation. legislation.gov.au. Accessed 11 October 2026.
  13. National Association of Testing Authorities, Australia (NATA). Accreditation programs, including ISO/IEC 17025 testing and calibration. nata.com.au. Accessed 11 October 2026.

Important information and disclaimer

General information only. This article is published for general educational and informational purposes. It summarises published scientific literature and public regulatory information as at the “last reviewed” date shown above. It is not medical, health, pharmaceutical, veterinary or legal advice, and it must not be relied on as a substitute for advice from a qualified professional about your circumstances.

Not for human or animal use. Nothing in this article recommends, endorses or describes the use of any substance in humans or animals. Findings from cell, animal or early human studies cannot be assumed to apply to people. The compounds discussed have not been evaluated by the Therapeutic Goods Administration (TGA) for safety, quality or efficacy unless the article expressly states otherwise. Any research materials offered by Iconic Peptides are supplied solely for in vitro and laboratory research.

No therapeutic claims. Iconic Peptides makes no claim that any product treats, cures, prevents, alleviates or diagnoses any disease, condition or symptom. References to published studies, clinical trials or medicines approved by regulators are factual descriptions of the scientific and regulatory record. They are not representations about any product we supply.

Regulatory status. Many peptides are scheduled substances under the Poisons Standard. Some are prescription-only (Schedule 4), and some may not be possessed without authority (Appendix D). The TGA has stated that a “research use only” label does not make supply lawful, change a product’s regulatory status, or remove advertising or supply obligations. State and territory laws may also apply. You are responsible for knowing and complying with all laws that apply to you. Regulatory information in this article may change after the review date.

Accuracy and third-party sources. We take care to cite and verify every source, but we do not warrant that this article is complete, current or error-free. Citing a study or organisation does not mean that its authors or that organisation endorse Iconic Peptides.

Limitation of liability. To the maximum extent permitted by law, Iconic Peptides, its owners, employees and contributors exclude all liability for any loss, damage, injury or expense, whether direct, indirect or consequential, arising from use of or reliance on this article or any misuse of any product. Nothing in this disclaimer excludes, restricts or modifies any right or remedy under the Australian Consumer Law that cannot lawfully be excluded.

Age restriction. This website is intended for adults aged 18 and over.

Certificate updates

Get notified when new certificates are published.

Cart

Your cart is empty.

Each product page shows the certificate status of every vial size.

Browse the catalogue
Subtotal
$0 AUD

Shipping is calculated at checkout.

Checkout

For laboratory and research use only. Not for human consumption. You declare your research purpose at checkout.

Before you enter

Iconic Peptides supplies analytically characterised materials for laboratory research only. They are not medicines, not for human or veterinary use, and not sold for any therapeutic purpose.

01
I am 18 years of age or older
02
I am visiting for research, educational or professional purposes
03
I understand nothing on this site is medical advice
Leave this site